Supplier of Alarm for Medical Gas Pipeline System in India|Supplier of Alarm for Medical Gas Pipeline System in India

Language
Spainsh German French Italian Dutch Danish English Japanese Korean
   
 
 

Our System & Devices

Medical Gas Pipeline System
Medical Gas Control & Automatic Change Over
Pipe Work for Medical Gas Distribution
Source Equipment for MGPS
Gas Distribution Network Control
Medical Outlet & Patient Interface Device
Anaesthetic Gas Scavenge Suction Systems
Digital Auto Change Over
Alarms for Medical Gas Pipeline Systems
Clean Room Modular OT
Hospital Equipment
Equipment Management System
Service & Support
Certificate
 
 

Send Enquiry

Name* :
Email* :
Mobile* :
Requirement :
 
   
 
 

Alarm for Medical Gas Pipeline System in India

 

nsp Hospitech offers two auto alarms to monitor current status of central medical gases supply & Vacuum system and trigger dual alarm in case of any deviation from the predetermined set parameters. These alarms alert the user with an auditory and a visual alarm for MGPS which can be conveniently located to any area / zone / operating rooms / ward areas / ICU/CCU in consultation with the health care personnel.

CRONUS Microprocessor Based Digital Alarm

 

CRONUS is a digital microprocessor based alarm system and sensing is done through highly sensitive transducer

URANUS Analog Alarm

URANUS is an analog Area Alarm which sense the set pressure deviations through pressure switch

 

Power Adapter
CRONUS has been designed for integration with the Master Alarm of central information systemwhich have 16 zone display facility. For detail refer to NSP Master Alarm literature. Although nominal pressure is pre-set in factory, but high & low alarm triggering range can be varied through re-set button in consultation with the health care personnel.

Both the Gas Alarm Systems can be ordered for any combination of medical gases including Vacuum and configured accordingly to a maximum of six gases for CRONUS and five gases for URANUS in the factory prior to shipment.

External power adapter that accepts 240 V AC (50 Hz) as the input voltage and delivers output voltage of 12V DC,

required only for auditory & visual signals.

The auditory signal can be cancelled only by the “MUTE” button or fault correction and once pressed the audio will remain mute for 15 minutes. However, the red display will blink and continue to indicate abnormality as long as a fault condition exists. Reset is automatic with the correction of the fault condition.

 

Medical Gas Pipeline Connection:

Mounting:

URANUS:

3/8” BSP Nut & Nipple

Wall only

CRONUS:

½” BSP Nut & Nipple

Wall & Conceal

Caution: To ensure that Gas specific and

Vacuum is connected to the Alarm ports

Dual Alarm :

Audio AlarmBuzzer- approx 70 dB
Visible SignalsGREEN Light “NORMAL”and RED “HIGH” or “LOW” condition.

Display Character : 0.5” red LED character, two figure for gases and three-figure for Vacuum

CRONUS Features :

Gas specific Temperature compensated silicon pressure sensor (Transducer). Operating temperature range - °C [°F] : -40 [-40] to 85 [185]
Humidity: (Dry gases ) : 0% to 95% RH, non-condensing Accuracy : ±0.25 % Full Scale Span
Total error band :  ±2 % Full Scale Span
Display Character : 0.5” red LED character, three-figure

High/Low set points - Set points shall be adjustable by two on board push buttons. “PUSH TO TEST” & “MUTE” buttons are easily accessible to operate and test the unit. Alarm have no moving parts and require no maintenance after initial installation.
Dust protected enclosure.

Option to keep Power Adapter external or park inside the unit. Option for Wall or concealed mounting
Provided with potential free contact (5 V DC) for integration with the Master Alarm

Technical Information:

Medical Gases

Nominal Pressure

Triggering of Alarm*

& Medical Services

ISO 7396 - 1:2007

Low

High

 

 

 

 

 

Oxygen, Nitrous Oxide,

58 psi

(4 bar)

Not < 58 psi (4 bar)

Not > 72 psi (5 bar)

Medical Air, Carbon Di-oxide

 

 

 

 

 

 

 

 

 

Air / Nitrogen for

116 psi

(8 bar)

Not <101 psi (7 bar)

Not >145psi(10 bar)

Surgical Tool

 

 

 

 

 

 

 

 

Vacuum

300 mm Hg  (5.8 psig)

Not < 260 mm Hg (5 psig)

- - - - -

 

 

 

 

 

* Factory settings.   Range can be altered in consultation with the health care personnel.

 

 

Ordering Information – Alarms

 

DESCRIPTION of

MEDICAL GASES &
VACUUM SERVICES

 

Code - CRONUS

Code - URANUS

2 GAS - OXYGEN  &  VACUUM

6062140

6062240

3 GAS - OXYGEN, MEDICAL AIR  & VACUUM

6063150

6063250

3 GAS - OXYGEN, NITROUS OXIDE  & VACUUM

6063160

6063260

4 GAS - OXYGEN, NITROUS OXIDE, MEDICAL AIR & VACUUM

6063170

6064270

5 GAS - OXYGEN, NITROUS OXIDE, MEDICAL AIR,AR 7  & VACUUM -

On Request

6065180

6 GAS - OXYGEN, NITROUS OXIDE, MEDICAL AIR,CO2, AR 7  & VACUUM -On Request

Not available

URANUS and CHRONUS have been designed to meet the Class IIa requirements of MDD 93 / 42 / EEC and amendment MDD / 2007 /47 / EC.

DISCLAIMER: nsp Alarms meet and exceed the requirements of relevant standard at the time of manufacture. However piping source, capacity, sizing and restrictions may prevent from attaining the stipulated value.

 
 
 
 
 
Download Brochure for more information
Our Location
Manufacturer India